tophat
Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
tophat [2020/06/29 09:54] – bpospich | tophat [2024/11/17 13:59] (current) – bpospich | ||
---|---|---|---|
Line 1: | Line 1: | ||
- | ====== HATPRO | + | ====== HATPRO-TOPHAT |
- | + | ||
- | {{: | + | |
- | // | + | |
- | + | ||
- | ===== Principle ===== | + | |
- | + | ||
- | **TOPHAT** is a passive microwave radiometer manufactured by [[https:// | + | |
- | + | ||
- | Zenith measurements alternate with RHI towards the north and PPI scans at 30 degrees elevation. Scan patterns are partially synchronized with the neighbouring cloud radar. Additionally, | + | |
- | + | ||
- | + | ||
- | ===== Instruments | + | |
- | + | ||
- | ^ Instrument name ^ current location | + | |
- | | {{: | + | |
- | | SUNHAT | + | |
- | | FOGHAT | + | |
+ | TOPHAT is a [[hatpro|HATPRO]] microwave radiometer, installed at Jülich. It was originally a G2 radiometer, and was upgraded in 2016 to G5 receivers. | ||
===== TOPHAT History ==== | ===== TOPHAT History ==== | ||
- | ^ Period | + | ^ Period |
- | | 22/07/2010 - today | [[http:// | + | | 24/07/2024 - 17/10/2024 | Cologne Biozentrum | | |
+ | | 22/07/2010 - 16/ | ||
+ | | 10/2016 | | Upgrade to G5 receivers | ||
| 07/08/2009 - 24/10/2009 | ARM AMF site , Cerro Toco, Chile [[http:// | | 07/08/2009 - 24/10/2009 | ARM AMF site , Cerro Toco, Chile [[http:// | ||
- | {{: | + | {{: |
// | // | ||
- | |||
- | ===== Measurement examples ===== | ||
- | |||
- | {{wiki: | ||
- | {{wiki: | ||
- | //Time series of brightness temperatures (left) and Integrated Water Vapor / Liquid Water Path (right)// | ||
- | |||
- | Current observations at [[http:// | ||
- | |||
- | |||
- | |||
- | |||
- | |||
- | |||
- | ===== Data ===== | ||
- | * If you are interested in receiving data please contact: B. Pospichal | ||
- | * [[http:// | ||
- | |||
- | |||
- | |||
- | The figure shows the relative humidity (RH, top row) and temperature (second row) of the environment as well as the Integrated Water Vapor (IWV, third row) and Liquid Water Path (LWP, last row) derived from TOPHAT. | ||
- | |||
- | |||
- | |||
- | ===== History ==== | ||
- | ^ Period | ||
- | | 22/07/2010 - today | [[http:// | ||
- | | 07/08/2009 - 24/10/2009 | ARM AMF site , Cerro Toco, Chile [[http:// | ||
- | |||
- | ===== Specifications ===== | ||
- | ^ Parameter | ||
- | | Receiver 1 (R1) | water vapor and liquid water absorption | ||
- | | Frequencies (R1) [GHz] | 22.24, 23.04, 23.84, 25.44, 26.24, 27.84, 31.40 | | ||
- | | Channel band-widths (R1) [MHz] | 230, 230, 230, 230, 230, 230, 230 | | ||
- | | Optical Resolution (R1) (HPBW) | ||
- | | Temporal Resolution | ||
- | | Receiver 2 (R2) | oxygen absorption | ||
- | | Frequencies (R2) [GHz] | 51.26, 52.28, 53.86, 54.94, 56.66, 57.30, 58.00 | | ||
- | | Channel band-widths (R2)[MHz] | ||
- | | Optical Resolution (R2) (HPBW) | ||
- | | Pointing Resolution (R2) | 0.6° (elevation), | ||
- | | Side-lobe level | < -30 dBc | | ||
- | | Infrared pyrometers | ||
- | | Size | 630 mm x 360 mm x 900 mm | | ||
- | | Power Consumption | ||
- | | Weight | ||
- | | Manufacturer | ||
- | |||
- | |||
- | Instrument time-line | ||
- | 07/08/2009 - 24/10/2009 ARM AMF site , Cerro Toco, Chile, RHUBC-II campaign | ||
- | 22/07/2010 - today Forschungzentrum Jülich (JOYCE-JuCol) | ||
- | 2016 Upgrade to HATPRO-G5 receivers | ||
- | Available measurement modes | ||
- | · Vertically pointing observations with integration times up to 1 seconds. | ||
- | · Elevation scans for boundary-layer temperature profiles. | ||
- | · Full azimuth/ | ||
- | JOYCE-CF Standard Operation Procedures | ||
- | · Standard measurement mode is vertically pointing, 1 second integration time | ||
- | · Every 15 minutes one elevation scan with following angles: 42°, 30°, 19°, 10.2°, 5.4° (duration ~100 seconds) | ||
- | · Every 30 minutes one azimuth scan at 30° elevation angle with 5° azimuth resolution (duration ~3 minutes) | ||
- | |||
- | · Automatic relative calibrations are routinely performed | ||
- | o automatic noise-switching for gain calibration with 50 Hz | ||
- | o System noise is calibrated every 5 minutes with view on ambient target | ||
- | · Absolute calibrations using LN2 are performed every 6 months | ||
- | Data quality assurance procedures | ||
- | · Automatic rain detection (data are flagged in case of rain detected by rain sensor) | ||
- | · Offset corrections for liquid water path | ||
- | · O-B statistics from atmospheric models for brightness temperature offset correction available | ||
- | · Spectral consistency checks | ||
- | · Thresholds for Level 2 data products | ||
- | · Housekeeping data analysis | ||
- | · Visual data inspection | ||
- | Available datasets | ||
- | The following data products are provided via the SAMD database. If you would like to have additional data or recent data that have not been uploaded to SAMD yet, please fill a data request sheet available at JOYCE-CF website. | ||
- | |||
- | Level 1 | ||
- | · Brightness temperatures (14 channels microwave) | ||
- | o Temporal resolution 1 second | ||
- | o 1 file per day (11-13 MB) | ||
- | o Filename: sups_joy_mwr00_l1_tb_v00_YYYYMMDDHHMMSS.nc | ||
- | |||
- | · Brightness temperatures (14 channels MW, 6 elevation angles) | ||
- | o Temporal resolution 15 minutes | ||
- | o 1 file per day (0.07 MB) | ||
- | o Filename: sups_joy_mwr00BL_l1_tb_v00_YYYYMMDDHHMMSS.nc | ||
- | |||
- | · Brightness temperature (2 channels infrared) | ||
- | o Temporal resolution 1 second | ||
- | o Filename: | ||
- | Level 2 | ||
- | · Integrated water vapor (IWV) | ||
- | o Temporal resolution 1 second | ||
- | o 1 file per day (3-4 MB) | ||
- | o Multivariate linear retrieval based on radiosoundings from De Bilt | ||
- | o Filename: sups_joy_mwr00_l2_prw_v00_YYYYMMDDHHMMSS.nc | ||
- | |||
- | · Integrated cloud liquid water (LWP) | ||
- | o Temporal resolution 1 second | ||
- | o 1 file per day (3-4 MB) | ||
- | o Multivariate linear retrieval based on radiosoundings from De Bilt | ||
- | o Filename: sups_joy_mwr00_l2_clwvi_v00_YYYYMMDDHHMMSS.nc | ||
- | |||
- | · Temperature profiles (from zenith pointing) | ||
- | o Temporal resolution: 1 second | ||
- | o 1 file per day (25-30 MB) | ||
- | o Multivariate linear retrieval based on radiosoundings from De Bilt | ||
- | o Filename: sups_joy_mwr00_l2_ta_v00_YYYYMMDDHHMMSS.nc | ||
- | |||
- | · Temperature profiles (from elevation scans for improved resolution in lowest 1000 m above ground) | ||
- | o Temporal resolution: 15 minutes | ||
- | o 1 file per day (0.04 MB) | ||
- | o Multivariate linear retrieval based on radiosoundings from De Bilt | ||
- | o Filename: sups_joy_mwrBL00_l2_ta_v00_YYYYMMDDHHMMSS.nc | ||
- | |||
- | |||
- | · Humidity profiles | ||
- | o Temporal resolution 1 second | ||
- | o 1 file per day (25-30 MB) | ||
- | o Multivariate linear retrieval based on radiosoundings from De Bilt | ||
- | o Filename: sups_joy_mwr00_l2_hua_v00_YYYYMMDDHHMMSS.nc | ||
- | Composite products using data from TOPHAT | ||
- | · Cloudnet target classification uses LWP data from TOPHAT. Combined with cloud radar and ceilometer observation, | ||
- | · Data are available via ACTRIS cloudnet database http:// | ||
tophat.1593417240.txt.gz · Last modified: 2020/06/29 09:54 by bpospich